OPERATING INSTRUCTIONS Humidity Chambers HCP 108 HCP 153 HCP 246
HCP page 2
1 General notes and safety notes............................................................................................................ 5 1.1 Transport ................................................................................................................................. 5 2 Installation .......................................................................................................................................... 6 2.1 Subframe (accessory) ..........................................
18 Communication interface for the PC ................................................................................................ 38 18.1 Communication interface RS232C ........................................................................................ 38 18.2 Bus interface RS485 ............................................................................................................. 39 19 Report memory ...........................................................................................
1 General notes and safety notes You have purchased a technically fully proven product which has been produced in Germany with the use of high-grade materials and the application of the latest manufacturing techniques; it has been factory tested for many hours. For this unit we guarantee availability of spares for 10 years Observation of the Operating Instructions is necessary forfaultless operation and for any possible claims under warranty.
2 Installation It is important that the oven is set up accurately horizontally; the door may have to be adjusted (see section 21) cm min. 20 The oven can be placed on the floor or on a bench (working surface). min. 8 min. 8 min. 15 cm cm cm The spacing from the back of the oven to the wall should be at least 15 cm. The spacing to the ceiling must not be less than 20 cm and that at the side to the wall not less than 8 cm. Generally it is essential to have adequate air ventilation around the oven.
2.3 Initial start-up When the oven is started up for the first time, it should be supervised continuously until steady conditions have been reached. Severe vibrations during transport may cause movement of the temperature probes in their holder inside the chamber. Note therefore that before the first start-up the temperature probes should be checked for their correct position and, if necessary, carefully aligned in their mounting (see ill). 2.
3 Technical data supply cable serial and parallel interface H2O mm (minimum height) inner glas door adjusting screw Model HCP 108 HCP 153 HCP 246 Chamber width A [mm] 560 480 640 Chamber height B [mm] 480 640 640 Chamber depth C [mm] 400 500 600 Housing width D [mm] 710 630 790 Housing height E (variable through adjustable feet) [mm] 778 920 938 Housing depth F (including door handle) [mm] 590 690 790 Chamber volume [litre] 108 153 246 Weight [kg] 70 90 110 Power [W] 1000 1500 2000 5 7 7 Max.
3.1 Working range of the HCP units The temperature-humidity diagram of the humidity chamber HCP indicates the range of temperature and humidity within which continuous condensation-free operation is possible. B Humidity %rh A C Temperature °C Range A: Within this range, any combination of temperature and humidity is possible without any appreciable condensation. Range B: Above the range indicated, e.g.
3.
3.3 Material quality For the external housing MEMMERT employs stainless steel (Mat.Ref. 1.4016). The chamber is made from stainless steel (Mat.Ref. 1.4301) which exhibits high stability, optimal hygienic properties and corrosion resistance against many (not all) chemical compounds (care required e.g. with chlorine compounds). The oven load has to be checked carefully for its chemical compatibility with the above materials. A compatibility table for materials can be requested from MEMMERT.
4 Oven construction and operation 1 2 1 1 1 1 3 The special feature of the HCP is the particularly uniform conditions inside the chamber. The air is heated by large-area allround heating (1). Humidity is provided by a dry steam generator which produces controlled evaporation of the water. The sterile dry steam is introduced into the chamber above the fan and mixed into the air flow (2). A continuous uniform atmosphere is ensured by the turbulence-free chamber ventilation (3). 4.
4.3 Switching on The oven is switched on by pressing the push/turn control. Oven switched off. The push/turn control is pushed in and protected against damage. Oven switched on and can be operated using the push/turn control and the SET key. 4.4 Setting the temperature (quick setting) Hold down the SET key and set the temperature setpoint with the push/turn control. After the SET key has been released the display briefly flashes the temperature setpoint.
7 Normal operation PRINT SETUP In this operating mode the oven operates continuously. The settings for operating the oven can be selected. The settings act directly on the operation of the oven.
Setting example Normal operation The unit has to heat up to 50°C at a humidity of 95%rh. The monitor function has to be activated at 51°C. temperature °C relative humidity (rh) % 100 100 80 80 60 60 temperature monitor 40 40 20 20 time time 1. Select operating mode “Normal operation“ After holding down the SET key (approx. 3 sec), the current operating mode is flashing. Select operating mode I with the push/turn control while holding down the SET key.
8 Weekly programmer PRINT SETUP In this operating mode the weekly programmer is activated and the oven switches on and off automatically at the programmed times. While the weekly programmer is in the OFF phase the unit is in stand-by mode. Heating and the supply of humidity are switched off, the controller display is dimmed and shows the clock time. The sequence of the weekly programmer is repeated every week. A maximum of 9 time blocks, each consisting of ON time and OFF time, can be programmed.
Programming example Weekly programmer The unit has to switch on at 9.30 hrs from Mo to Fr (workday group) and switch off at 19.00 hrs. In addition it has to operate on Saturday from 10.00 to 14.00 hrs. 9:30h - 19:00h Fr Su 9:30h - 19:00h Th 9:30h 14:00h 9:30h - 19:00h We Sa 9:30h - 19:00h Tu 01:00h 9:30h - 19:00h 12:00h Mo 24:00h 1. Select operating mode “Weekly programmer“ After holding down the SET key (approx. 3 sec) the current operating mode is flashing.
9 Programme operation SETUP PRINT In this operating mode it is possible to set up to 40 freely programmable ramps for temperature and humidity.
Mo Tu We Th Fr Sa on 4 t4 t2 h off t3 Su t1 STERI loop IN 1 DEFRO °C 3 2 1 OUT %rh °C mb MAX MIN AUTO SETUP PRINT IN 2 Duration of first ramp segment Range: 1 minute to 999 hours.
9.1 Closure commands for ramp segments NEXT Follow-on with next programme segment. SPWT (T) Wait until the setpoint temperature is reached. The unit only starts the next programme segment when the programmed setpoint temperature has been reached, even if the programmed heating SET-POINT WAIT (temperature) time has already elapsed. SPWT (H) Wait until set humidity is reached.
Programming example programme operation On Monday at 08.00 hrs the unit has to heat up as quickly as possible to 50°C and reach a relative humidity of 95%rh. Once the temperature and humidity have been reached, the unit has to hold the setpoints for 45 minutes and then cool down to 37°C at a humidity of 50%rh. °C wait until set temperature and set humidity have been reached 37°C Mo 8.00 h 50°C 0.01 h 0.45 h 1.00 h t sufficient if SPWT (TH) is set as segment closure command 1.
6. Select temperature of first ramp segment Turn the push/turn control clockwise until the temperature display is flashing. Hold down the SET key and set the required temperature setpoint of 50.0 °C using the push/turn control. 7. Select relative humidity of first ramp segment Turn the push/turn control clockwise until the humidity display is flashing. Hold down the SET key and set the required humidity setpoint of 95.0 %rh using the push/turn control. °C %rh 8.
14. Select temperature of third ramp segment Turn the push/turn control clockwise until the temperature display is flashing. Hold down the SET key and set the required temperature setpoint of 37.0 °C using the push/turn control. °C 15. Select relative humidity of third ramp segment Turn the push/turn control clockwise until the humidity display is flashing. Hold down the SET key and set the required humidity setpoint of 50.0 %rh using the push/turn control. %rh 16.
10 Printer PRINT SETUP All HCP ovens are fitted as standard with a parallel printer interface, as used on personal computers. This parallel printer interface on the back of the oven is suitable for connecting conventional PCL3compatible printers which are provided with a parallel printer interface (e.g. HP Deskjet 5550 or HP Deskjet 9xx). It is important to use a screened interface cable. The screen must be connected to the plug case.
11 Basic oven settings (Setup) PRINT SETUP In this operating mode it is possible to make the basic settings of the oven. Clock time, date, day, year, and settings of sounder, of address assignment, monitoring units, heater power and calibration are set here.
Calibration correction for user-calibration (see section 17) Exit setup mode = store all settings and exit SETUP mode 11.1 Real-time clock The real-time clock is set in SETUP and includes date and clock time. The real-time clock serves for documentation according to GLP. Date and clock time are marked in the report print. On the graphics print the time axis is marked in real-time. The clock runs with a buffer battery independently of the mains power supply.
12.2 Electronic temperature monitor 12.2.1 Overtemperature protection Adjustment range: up to 10°C max above nominal temperature (for nominal temperature see label) °C MAX MIN Using the push/turn control select the symbol MAX . Hold down the SET key and set the protection temperature using the push/turn control. AUTO 12.2.2 Undertemperature protection Adjustment range: from 10°C below minimum temperature of oven to 10°C above nominal temperature of oven (for nominal temperature see label).
12.2.4 Automatic temperature monitor (ASF) A monitoring device which automatically follows the selected temperature setpoint. Automatic temperature monitor OFF °C (ASF OFF) Using the push/turn control select the symbol AUTO . Hold down the SET key and select off using the push/turn control. MAX MIN AUTO Automatic temperature monitor ON °C (ASF ON) Using the push/turn control select the symbol AUTO . Hold down the SET key and select on using the push/turn control.
Going outside tolerance margin = ASF alarm Going outside the selected tolerance margin of the setpoint (in the example 37°C, +/-3°C), for example through opening the oven door during operation, triggers the alarm. Triggering the ASF alarm is indicated by flashing symbol AUTO and . If the sounder is switched on in SETUP, the ASF alarm is additionally signalled by an interrupted tone. By pressing the SET key the sounder can be switched off temporarily until the next occurrence of an alarm event.
Setpoint changed = ASF de-activated automatically If the temperature setpoint is altered, the automatic temperature monitor is automatically de-activated temporarily (see in the example the setpoint is changed from 37°C to 30°C) until the tolerance margin of the new temperature setpoint is reached (see in the example below: the ASF is re-activated at 30°C, +/3°C).
14 Audible alert signals The interval sound indicates a fault in the temperature control system or the humidifying system. It is triggered in the following cases: Fault in the temperature control system: Tb activ - temperature limiter activated if the temperature limiter operates (see section 12.1) hi-alarm - overtemperature alarm activated if the overtemperature protection operates (see section 12.2.3) lo-alarm- undertemperature alarm activated if the undertemperature protection operates (see section 12.2.
15 Sterilisation chip card Sterilisation Parameter: 160 °C / 4:00 h Unit-ID: ________________________ Sterilisation Process Control Card As protection against contamination the HCP is equipped as standard with a STERICard. The STERICard starts an automatic and pre-set sterilisation procedure. To activate, insert the STERICard into the slit in the control field, select the Start symbol turn control and the depressed SET key.
16 User-ID-Card (available as optional extra) Name: _____________________ ID: _____________________ access authority card The User-ID-Card stores the serial number of the oven and a unique user number in encrypted format. The User-ID-Card therefore functions only in the oven with the corresponding serial number. Each log-on via the User-ID-Card is documented in the internal flash memory. If the USER-IDcard is inserted, the SETUP menu includes the additional item ID-LOCK.
17 Calibration 17.1 Temperature User-calibration of oven at the controller, with three calibration temperatures selected by the user. CAL.1 Temperature calibration at low temperature CAL.2 Temperature calibration at medium temperature CAL.3 Temperature calibration at high temperature Either a positive or a negative calibration correction can be applied to each selected calibration point. General calibration instructions: 1. 2. 3. 4. 5.
calibration point1 °C calibration point 2 °C calibration correction adjustment range -4.9 °C to +4.9 °C calibration temperature °C calibration point 3 calibration correction adjustment range -4.9 °C to +4.9 °C calibration temperature °C calibration correction adjustment range -4.9 °C to +4.9 °C calibration temperature °C °C Note: If all calibration corrections are set to 0.0 °C the factory temperature calibration is restored.
17.2 Humidity User-calibration of the unit at the controller, using two humidity points: RH 20 Humidity calibration at 20% relative humidity RH 90 Humidity calibration at 90% relative humidity Either a positive or a negative calibration correction can be applied to each selected calibration point. General calibration instructions: 1. Select the required calibration humidity in SETUP and set the corresponding calibration correction to 0.0 °C. 2.
calibration humidity 20%rh Mo Tu We Th Fr Sa calibration correction adjustment range -5.0 % to +5.0 % t3 Su on 4 t4 t2 t1 off STERI loop IN 1 DEFRO °C 3 2 1 Tu We Th Fr Sa IN 2 OUT CO2 mb mb AUTO calibration correction adjustment range -5.0 % to +5.
18 Communication interface for the PC 18.1 Communication interface RS232C The oven is provided as standard with a serial communication interface RS232C according to DIN 12 9001. Using this interface it is possible to control the oven from the PC and to produce reports. This is done using the “Celsius 2007“ software. For this purpose the oven has to be assigned a unique device address in sub-menu SETUP, option ADDRESS; This is the address through which the PC communicates with the oven.
18.2 Bus interface RS485 When so ordered, the oven can be equipped at the factory with an RS485 interface instead of the RS232C interface. This permits networking of several ovens (up to 16) with a single PC using a common 2-wire circuit. The system is operated using the “Celsius 2007” software. A unique device address has to be assigned to the oven in sub-menu SETUP, option ADDRESS. This is the address through which the PC communicates with the oven. The default setting is ADDRess 0.
19 Report memory The controller continuously records all relevant measurements, settings and error messages at 1-minute intervals. The internal report memory is arranged as a ring memory, i.e. the new data always overwrite the oldest report data. The report function can not be switched off but remains active at all times. The data are stored in the controller, protected against any manipulation. The controller memory can be read to produce documentation. Every data set is stored with a unique date stamp.
20 Cleaning By regular cleaning of the (easy to clean) interior, residues are avoided which may impair the outfit and function of the incubator. The unit can be cleaned with commercial stainless steel cleaning agent. Please note that objects liable to rust must not be placed into the interior. Rust sediments lead to contamination of the interior or the external casing. If rust stains caused by contamination occur on the interior surface, the affected areas must be cleaned and polished immediately.
21 Maintenance Important for a long life of your MEMMERT product and in case of warranty claims. Note: Any work involving opening up the oven must only be carried out by a suitably qualified electrician! MEMMERT products require little maintenance. It is however recommended to lubricate all moving parts of the doors (hinges and closure) once a year (or 4 times a year with continuous operation) using a thin Silicone grease, and to check that the hinge screws are tight.
22 Check list for fault rectification Mains switch switched on: • no indication on the display of the temperature module • no fan operation Symbol nor alight Symbol is alight Symbol is flashing • and rh empty • and rh over E-0 in temperature module E-1 in temperature module E-2 in temperature module E-3 in temperature module E-6 in humidity module error sensor rh E-L1 in timer module equipment fuse has blown; 15A instrument fuse has blown; T100mA 250V~ on board 55162.x controller faulty (board 55616.
Supply failure in operating mode “Normal operation“ After a supply failure the operation is continued with the set parameters. The instant and duration of the supply failure are documented in the record memory. Supply failure in operating mode “Weekly programmer“ After a supply failure the operation is continued with the set parameters. The instant and duration of the supply failure are documented in the record memory.
24 CE Conformity Declaration EC Declaration of Conformity Manufacturer´s name and address: MEMMERT GmbH + Co. KG Äußere Rittersbacher Straße 38 D-91126 Schwabach Humidity Chambers HCP 108 l / 153 l / 246 l AC 230 V 50/60 Hz Product: Type: Sizes: Nominal voltage: The designated product is in conformity with the European EMC-Directive 2004/108/EEC including amendments Council Directive of 03 May 1989 on the approximation of the laws of the Member States relating to electromagnetic compatibility.
Standard ovens (HCP) are safety-approved and bear the test marks: This product is subject to the Directive 2002/96/EC by the European Parliament and the EU Council of Ministers which concerns Waste Electrical and Electronic Equipment (WEEE). This product has been put on the market after 13 August 2005 in countries which have already incorporated this Directive into National Law. It should not be disposed off as part of domestic refuse. For disposal please contact your dealer or the manufacturer.
page 47 HCP
Memmert GmbH + Co. KG | Postfach 1720 | D-91107 Schwabach | Tel. +49 (0) 9122 / 925 - 0 | Fax +49 (0) 9122 / 145 85 | E-Mail: sales@memmert.com | www.memmert.com 28.10.