Instruction Manual

46
CELSIUS 10 FDA-Edition
8.2.3
Saving and loading protocol data (*.
pro)
If the option "
Create Log File" is enabled when a tempering profile is run (see page
35
), the
protocol values are saved to the
protocol file (*.
pro) selected. You may also write the actual
value curve with the corresponding tempering profile later on to a
protocol file if the protocol
option is not enabled in the
start dialog of the temperature programme.
The
protocol can only be saved
later for as long as it is displayed
in the
graphics window.
To create a file for the
protocol dis-
played, select "File"
"
Save Protocol
Data"
"To File (*.
pro)". In the win-
dow appearing, you can select
the
protocol file (button "File") and GLP
data that are to be recorded. This
option is automatically enabed if GLP
data have been set (see page
9
).
Protocol data saved in this way can
be re-
opened
with "File"
"
Load Pro-
tocol Data"
"From File" or
Load Pro-
The unique device number of the
connected MEMMERT equipment
is read by CELSIUS FDA and stored
with each protocol, with protection against alteration.
In order to prevent any alteration of files, CELSIUS FDA encrypts all profile and protocol
files. Encryption employs a key which is specified in the
licence file. Existing files from a
previous CELSIUS version can not be read or exported. If an encrypted file has been altered
(except through CELSIUS), access to the data is no longer possible, even for the producer
of the software.
8.2.4
Loading protocol data from the internal MEMoryCard of the
oven
If the oven is equipped with a MEMor
yCard, protocol data are stored on it. They can be read
If the oven is equipped with a MEMoryCard, protocol data are stored on it. They can be read If the oven is equipped with a MEMor
in with CELSIUS
via
"File"
"
Load Protocol Data"
"From Internal MemoryCard
" or
.
8.2.5
Loading protocol data (*.bin) from the
ring protocol memory of the oven
The oven saves all values measured over a long period automatically in its
internal ring
protocol memory. These measured values can be read in with CELSIUS
("File"
"
Load Protocol
Data"
"From Device
(
"From Device ("From Device
Ring Protocol Memory)" or
).
).
When they are first read out, the entire ring protocol memory is read out and saved as a file in
the format *.bin. This file is write-protected. The preset directory for ring protocol files varies
according to the operating system. This can be displayed via "Settings"
→"
Open Common CEL-
SIUS Directory". The ring protocol data are located in the subdirectory of this, "FlashProtocols".
Reading out the protocol memory for the first time may take some time (up to 15 min-
utes), depending on the device runtime.