AtmoCONTROL
Depiction
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
?
15
STAND BY
15
Hold pres-
sure
+
-
Tolerance band
Alarm
Safe
on off
hold p
20mb
10mb
2h:30 m
500mb
Function
Maintains a specific pressure for a
specific time.
Adjustment options
Adjustment options
Name of programme segment
1
Duration (time or infinite
)
Pressure to be maintained
Tolerance value above/below
Alarm if limits are exceeded
Safe
2
?
15
STAND BY
15
Change
pressure
+
-
Tolerance band
1100mb
decrease
2h:30 m
500mb
10mb
20mb
SPWT
on off
Function
Increases or decreases pressure
over a specific time to a specific
value.
Adjustment options
Adjustment options
Name of programme segment
1
Duration
Target (setpoint) pressure
Tolerance value above/below
SPWT
3
1
When run, this is displayed in the
status bar of the appliance
2
When
Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
3
SPWT:
Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.