Manual

16
D24042 | Date 09/2014
AtmoCONTROL
4.1.4
Available
parameters
Below, all the parameter icons with their adjustment options are shown.
Which parameters are available to adjust the programme depends on the appliance for
which a programme is to be created. Only those parameters are available that the appli-
ance is able to implement. For appliances without humidity regulation, for example, no
humidity icon is available. The respective adjustment options (temperature ranges etc.) are
appliance-specific.
Broad
parameter representation
parameter representation
Depiction
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
?
15
STAND BY
15
Hold tem-
perature
Tolerance band
Function
Maintains a set temperature for a
specific time.
Adjustment options
Adjustment options
Name of programme segment
1
Duration (time or infinite
)
Temperature to be maintained
Tolerance value above/below
Alarm if limits are exceeded
Safe
2
(For programme example, see page
24)
?
15
STAND BY
15
Change
tempera-
ture
0,1 K/min
Tolerance band
Function
Increases or decreases temperature
over a specific time to a set value.
Adjustment options
Adjustment options
Name of programme segment
1
Duration
Target (setpoint) temperature
Tolerance value above/below
SPWT
3
(For programme example, see page
25)
1
When run, this is displayed in the
status bar of the appliance
2
When
Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
3
SPWT:
Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.