User Manual

D30396 | Date 09/2014 21
AtmoCONTROL FDA Version
Narrow
parameter representation
parameter representation
With the narrow
parameter representation, no time progression can be set, in contrast to
broad parameter representation. The setting made immediately becomes active at the respec-
tive position – and remains active until it is changed by the insertion of a new parameter icon
of the same type.
Depiction
in icon bar
Meaning Depiction on
editor thread
Adjustment options/
comments
?
15
STAND BY
15
CO
2
CO
2
15
0
50%
0 to 20 percent
For a setpoint ≠ 0.0, the fan is automati-
cally set to 50 %.
?
15
STAND BY
15
O
2
CO2
12.0%
O2
1 to 20 percent, off
?
15
STAND BY
15
Fan speed
100%
10
0
%
FAN
0 to 100 percent in steps of 10 %
(For programme example, see page 28)
?
15
STAND BY
15
Air flap
position
100%
FLAP
0 % (closed, recirculating operation) to 100 %
(opened completely, fresh air operation) in
steps of 10%
(For programme example, see page 28)
?
15
STAND BY
15
Interior
lighting
100%
LIGHT
depends on appliance type
0 or 100 % (off/on)
0 to 100 % in steps of 1%
?
15
STAND BY
15
UV light
UV LIGHT
on off
on/off
?
15
STAND BY
15
Horn
HORN
Adjustment options: none
Appliance emits an acoustic signal at the
position in the programme at which the icon
was inserted, for example if a specific setpoint
value is reached or the programme is finished.