User Manual

D30396 | Date 09/2014 19
AtmoCONTROL FDA Version
Broad
parameter representation
parameter representation
Depiction in
icon bar
Mean-
ing
Depiction on
editor thread
Function and
adjustment options
?
15
STAND BY
15
Hold
temper-
ature
Tolerance band
Function
Maintains a set temperature for a
specific time.
Adjustment options
Adjustment options
Name of programme segment
1
Duration (time or infinite
)
Temperature to be maintained
Tolerance value above/below
Alarm if limits are exceeded
Safe
2
(For programme example, see page
26)
?
15
STAND BY
15
Change
temper-
ature
0,1 K/min
Tolerance band
Function
Increases or decreases temperature
over a specific time to a set value.
Adjustment options
Adjustment options
Name of programme segment
1
Duration
Target (setpoint) temperature
Tolerance value above/below
SPWT
3
(For programme example, see page
27)
?
15
STAND BY
15
Hold
humid-
ity
5.0%rh
Tolerance band
+
-
Alarm
Safe
on off
70.0%rh
12h:00m
Hold
0.0%rh
Function
Maintains a specific humidity for a
specific time.
Adjustment options
Adjustment options
Name of programme segment
1
Duration (time or infinite
)
Humidity value to be main-
tained
Tolerance value above/below
Alarm if limits are exceeded
Safe
2
1
When run, this is displayed in the
status bar of the appliance
2
When
Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
3
SPWT:
Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.