ADDITIONAL ACCESSORIES 100% ATMOSAFE. MADE IN GERMANY. www.memmert.com | www.atmosafe .
Manufacturer and customer service Memmert GmbH + Co. KG Willi-Memmert-Straße 90–96 D-91186 Büchenbach Germany Phone: +49 (0)9122 925-0 Fax: +49 (0)9122 14585 E-mail: sales@memmert.com Internet: www.memmert.com Customer service: Service hotline: +49 (0)9171 9792 911 Service fax: +49 (0)9171 9792 979 E-mail: service@memmert.com Shipping address for repairs: Memmert GmbH + Co. KG Customer service Willi-Memmert-Str.
About this manual About this manual Purpose and target audience This manual describes the optional additional accessories for Memmert Generation 2012 appliances. It is intended for use by trained personnel of the owner, who have the task of operating and/or maintaining the respective appliance.
Contents Contents Floating switching contact ALARM (opt. H6) Floating switching contact setpoint reached (SP) (opt.
Additional accessories Floating switching contact ALARM (opt. H6) Relay/ LED Relay off Test LED red Relay on Test LED green Plug assignment Functional description Miscellaneous Contact 2-3 is closed in case of the following errors: ► Loss of voltage ► Overtemperature ► Undertemperature (plus controller only) ► Humidity alarm ► Mechanical Temperature limiter TB ► Fan speed alarm (only for optional fan speed monitoring) ► Error of sensor PT100 Switching capacity max. 2 A max.
Additional accessories Floating switching contact setpoint reached (SP) (opt. H5) Relay/ LED Relay off Test LED red Relay on Test LED green 6 Plug assignment Functional description Miscellaneous Contact 2-3 is closed ► Setpoint not reached ► The actual temperature value is outside of a set tolerance band around the defined setpoint Switching capacity max. 2 A max. 24 Volt Appliance type Tolerance band IN, IF, IPP, HPP, INCO2, HCP, ICP, ICH, IPS dT ≥ 0.5 K UN, UF, SN, SF, CTC, TTC dT ≥ 2.
Additional accessories Freely programmable switching contact (A – D) Relay/ LED Relay off Test LED red Relay on Test LED green Plug assignment Functional description Miscellaneous Programming of the freely programmable switching contacts is done using the AtmoCONTROL software. Up to 4 (for singlephase appliances max. 2) floating switching contacts can be switched programme dependently. For switch setting "open" ► Contact 1-2 open ► Contact 2-3 closed Switching capacity max. 2 A max.
Additional accessories 4–20-mA current loop interface for temperature Plug assignment Appliance Range 4 mA 12 mA 20 mA IN / IF 0 … + 90 °C 0 °C 45 °C 90 °C IN /IF (with Steri function) 0 … + 90 °C 0 °C 45 °C 90 °C SN / SF 0 … + 260 °C 0 °C 130 °C 260 °C UN / UF 0 … + 310 °C 0 °C 155 °C 310 °C IPP / HPP –10 … + 80 °C – 10 °C 35 °C 80 °C ICP / ICH –20 … + 70 °C – 20 °C 25 °C 70 °C CTC –50 … + 200 °C – 50 °C 75 °C 200 °C Optional 0 … + 70 °C 0 °C 35 °C 70 °C Optio
Additional accessories Electric door locking mechanism Functional description Programming of the electric door locking mechanism is done using the AtmoCONTROL software. You can set the door locking mechanism to be electrically locked and unlocked at any stage of the programme.
Additional accessories HPP constant climate chamber with compressed air dehumidification In this special variation, the temperature-humidity working area of the appliance is extended by the use of external compressed air.
Notes 11
27.02.2014 D24124 // englisch Memmert GmbH + Co. KG Willi-Memmert-Straße 90-96 | D-91186 Büchenbach Tel. +49 9122 925-0 | Fax +49 9122 14585 E-Mail: sales@memmert.com HPP/IPPplus deutsch facebook.com/memmert.family Die Experten-Plattform: www.atmosafe.net 15.06.