8.4.2 Differences with the standard SMBus specification (reference [1])
There are eleven command protocols for standard SMBus interface. The MLX90614 supports only two of
them. Not supported commands are:
• Quick Command
• Byte commands - Sent Byte, Receive Byte, Write Byte and Read Byte
• Process Call
• Block commands – Block Write and Write-Block Read Process Call
Supported commands are:
• Read Word
• Write Word
8.4.3 Detailed description
The PWM/SDA pin of MLX90614 can operate also as PWM output, depending on the EEPROM settings. If
PWM is enabled, after POR the PWM/SDA pin is directly configured as PWM output. The PWM mode can be
avoided and the pin can be restored to its Data function by a special command. That is why hereafter both
modes are treated separately.
8.4.3.1 Bus Protocol
Figure 4: SMBus packet element key
1181 17 1
S
WrSlave Address A Data Byte
Slave-to-Master
Master-to-Slave
Packet Error Code
PEC
Stop Condition
S
Acknowledge (this bit can be 0 for ACK and 1 for NACK)
A
Write (bit value of 0)
Wr
Read (bit value of 1)
Rd
Repeated Start ConditionSr
Start Condition
S
AP
After every 8 bits received by the SD an ACK/NACK takes place. When a MD initiates communication, it first
sends the address of the slave and only the SD which recognizes the address will ACK, the rest will remain
silent. In case the SD NACKs one of the bytes, the MD should stop the communication and repeat the
message. A NACK could be received after the PEC. This means that there is an error in the received
message and the MD should try sending the message again. The PEC calculation includes all bits except the