User Manual
RedLab 1208LS User's Guide  Functional Details 
21 
With all outputs at their maximum output current, you can calculate the total current requirement of the RedLab 
1208LS USB +5 V as follows: 
(RedLab 1208LS @ 20 mA) + (16 DIO @ 2.5 mA ea) + (2 AO @ 30 mA ea ) = 120 mA 
For an application running on a PC or powered hub, the maximum available excess current is 500 mA−120 mA 
= 380 mA. This number is the total maximum available current at the PC+5V screw terminals. Measurement 
Computing highly recommends that you figure in a safety factor of 20% below this maximum current loading 
for your applications. A conservative, safe user maximum in this case would be in the 300-320 mA range. 
Since laptop computers typically allow up to 100 mA, the RedLab 1208LS in a fully-loaded configuration may 
be above that allowed by the computer. In this case, you must determine the per-pin loading in the application 
to ensure that the maximum loading criteria is met. The per-pin loading is calculated by simply dividing the 
+5 V by the load impedance of the pin in question. 
Ground terminals 
The 10 ground (GND) connections are identical, and provide a common ground for all RedLab 1208LS 
functions. Refer to the pinout diagrams
 on page 16 for the location of the GND terminal pins. 
Calibration terminal 
The CAL connection (pin 16) is an output you should use only to calibrate the RedLab 1208LS. Refer to the 
pinout diagrams on page 16 for the location of this pin. Calibration of the RedLab 1208LS is software-
controlled via InstaCal. 
Trigger terminal 
The TRIG_IN connection (pin 18) is an external digital trigger input. You can configure this terminal with 
software for either trigger high or trigger low. 
Counter terminal 
Pin 20 (CTR) is input to the 32-bit external event. Refer to the pinout diagrams on page 16 for the location of 
this pin. The internal counter increments when the TTL levels transition from low to high. The counter can 
count frequencies of up to 1 MHz. 
Accuracy 
The overall accuracy of any instrument is limited by the error components within the system. Quite often, 
resolution is incorrectly used to quantify the performance of a measurement product. While "12-bits" or "1 part 
in 4096" does indicate what can be resolved, it provides little insight into the quality of an absolute 
measurement. Accuracy specifications describe the actual results that can be realized with a measurement 
device. 
There are three types of errors which affect the accuracy of a measurement system: 
  offset 
  gain 
  nonlinearity. 
The primary error sources in the RedLab 1208LS are offset and gain. Nonlinearity is small in the RedLab 
1208LS, and is not significant as an error source with respect to offset and gain. 










