Specifications
S.2Transmitted or
Incident Light
does
not work
1 . Make
sure
your
mains
outlet
indeed has
power.
2.
Check
to
see
if
the
mains fuse
has blown.
3. Check
to
see
if the lamp has blown.
5.3
Replacing Transmitted
6V
30W
Halogen Lamp
CAUTION
ALWAYS DISCONNECT YOUR EQUIPMENT
BEFORE DOING ANY REPAIR.
Do not install
any bulb
in
your
instrument
other than ones designated by
Meiji
Techno:
For
All Integrated
llluminators MT920O, MT9300, MT9420
&
MT9430
I
Do not touch the
glass
envelope of
the lamp
during
installation. Keep the
protective
sleeve or bag of the
lamp during installalion and
remove
it
right
after installation.
l@HdTt-l
LAM' AND LAM' HousrNG MAy BE Hor ro rHE roucH.
o
Switch
otf
the
microscope.
o
Disconnect the
power
cord.
o
Wait until the
housing
and
bulb have
cooled
sufficiently.
o
Remove the illuminator
cover screws and
lift
the
cover to
remove.
o
Replace
the defective
lamp.
o
Place
a
new lamp into the socket while
avoiding
touching the
glass.
Notice the
bulb
is
wrapped in
plastic.
Use
it to
avoid
touching the
glass
envelope
during
installation.
o
Be
sure the bulb
is
pushed
in
as
far
as
possible
thereby
aligning
the filament
in the illuminator.
o
Reinstall the lamp
cover door.
o
Reconnect
the
power
cord
and
switch on
the
microscope
to
verify
proper
illuminator
operation.
6V 30W
Halogen
25