Datasheet
SALES OFFICE
Megger Limited
Archcliffe Road Dover
CT17 9EN England
T +44 (0) 1304 502101
E UKsales@megger.com
MIT4002_DS_en_V01
www.megger.com
ISO 9001:2008
The word ‘Megger’ is a registered trademark
ORDERING INFORMATION
Description Order Code
MIT400 - 250V, 500V 1000V Insulation tester 1004-731
MIT410 - as MIT400 + 50 V & 1000 V, PI, DAR
and timed testing 1004-735
MIT420 - as MIT410 + variable voltage test,
storage & recall 1004-739
MIT405 1004-734
MIT415 - 10 V, 25 V, 50 V, 100 < 250 V, 500 V I
nsulation tester 1006-738
Included Accessories
Red/Black Silicone test leads with Probes + Clips
SP5 remote switch probe (not MIT400)
Owners information CD
Batteries 6 x AA Alkaline (MIT420 & 430 MiMH)
Hard carry case
Description Order Code
Optional Accessories
UKAS calibration certificate
Mains Charger kit (MIT420 & 430 only) 1007-464
DC battery charger 1004-183
Switched probe SP5 1007-157
Test lead set and crocodile clips 1002-001
2 wire 500 mA fused test lead set 1002-015
Batteries (6 x NiMH) 1002-753
Weight (instrument and
case)
1.75 kg (3.86 lb)
Fuse Use only 2 x 500 mA (FF) 1000 V
32 x 6 mm ceramic fuse of high
breaking capacity HBC 10 kA
minimum. Glass fuses MUST NOT
be fitted.
Safety protection The instruments meet EN 61010-
1 (1995) to 600V phase to earth,
Category IV. Refer to safety
warnings supplied.
EMC In accordance with IEC 61326
including amendment No.1
Temperature co-efficient <0,1% per °C up to 1 GΩ
<0,1% per °C per GΩ above 1
GΩ
Environmental:
Operating temperature
range and humidity
-10 to +55 °C
90% RH at 40 °C max.
Storage temperature range -25 to +70 °C
Calibration temperature +20 °C
Maximum altitude 2000 m
IP rating IP 54