Datasheet
LTW315 LTW325 LTW335 LTW425
Features
Backlit display
(White light) n n n n
Built-in front cover
(Foldaway) n n n n
IP54 Weatherproof case n n n n
Accepts rechargeable
batteries n n n n
Test result storage n
Downloading test results n
Test lead set with crocs n n n n
Mains socket test lead n n n n
Full Calibration Certificate n n n n
IEC61010-1 300V CATIV n n n n
EN61557 n n n n
Specification
General Specification
Only values with tolerances or limits are guaranteed data. Values
without tolerances are for information only.
Accuracy
All accuracy statements are based on:
Ambient temperature: 23º ± 2ºC
Nominal source voltage: 230 V a.c ± 1%
Voltage measurement
(ac only) 50V to 440 V
Accuracy: ±2% ±1V
Frequency measurement
Range: 25Hz to 99.99Hz
Accuracy: ±0.1Hz
A warning will be shown if applied voltage exceeds 440 V.
Damage will occur if applied voltage exceeds 600 √ rms.
Loop testing
Accuracy
+/-5% +/-0.03 ohms @ 230 V a.c.
+/-10% +/-0.03 ohms @ 100 V a.c. and 300 V a.c. (LTW325 &
LTW335)
+/-10% +/-0.03 ohms @ 100 V a.c. and 280V (LTW315)
+/-10% +/-0.02 ohms (LTW425)
+/-15% +/-0.03 ohms @ 50 V a.c. (LTW325 & LTW335)
+/-15% +/- 0.03 ohms @ 50 V a.c. (LTW425)
NO-TRIP Loop and High current Loop test
Source voltage:
LTW315 100V to 280V (49Hz to 50.1Hz)
LTW325,335 & 425 50V to 440V (49Hz to 50.1Hz)
Display range: 0.01 Ω to 2000 Ω
Accuracy: ±5% ±0.03 Ω
Nominal test current:
No-Trip loop test 15 mA (at nominal 230 V a.c supply)
High current loop test 4 A (at nominal 230 V a.c supply)
EN61557
Operational range: 0.30 Ω to 1000 Ω
High resolution loop test (LTW425 only)
Source voltage: 50 V to 440 V (50Hz)
Display range: 0.001 Ω to 2000 Ω
Accuracy: ±5% ±0.01 Ω
Nominal test current: 4 A at nominal 230 V supply
EN61557
Operational range: 0.30 Ω to 1000 Ω
Prospective Fault Current (PFC)
Prospective fault current = Measured Source Voltage / Loop r
esistance
Maximum range: No trip mode: 20 KA
High current mode: 20 KA
Hi resolution mode: 40 KA
Accuracy is derived from loop test and voltage measurement.
Environmental
Temperature and humidity
Operating Range: -10°C to +50°C
Operating Humidity: 90% R.H. non-condensing at
+40°C max.
Storage Range: -25°C to +70°C
Maximum altitude: 2000m to full safety specification
Dust and water ingress: IP54
Safety
IEC61010
Designed to IEC61010-1
Designed for 300 volts to Earth Category IV, with Phase to Phase
voltages to 440 Volts.
Fuse protected to 600 volts rms ac.
EN61557
Complies with the following parts of EN61557, Electrical safety in
low voltage systems up to 1000V ac and 1500V dc- Equipment for
testing, measuring or monitoring of protective measures:
Part1-General Requirements
Part3-Loop resistance
E.M.C
In accordance with IEC61326-1
Power supply
Battery: 8 x 1,5 V cells IEC LR6 type (AA alkaline).
Rechargeable: 8 x 1.2VNiCd or NiMH cells
Battery life: 2000 consecutive tests
LTW300