19.  Specifications
19.1  Accuracy
19.1.1.  Insulation test
Specfication Detail
1000 Volts  10 kΩ - 999 MΩ ±3% ±2 digits
500 Volts.  10 kΩ - 500 MΩ ±3% ±2 digits >500 MΩ ±10% ±4 digits
250 Volts.  10 kΩ - 250 MΩ ±3% ±2 digits >250 MΩ ±10% ±4 digits
100 Volts.  10 kΩ - 100 MΩ ±3% ±2 digits >100 MΩ ±10% ±4 digits
EN61557 range:  10 kΩ - 999 MΩ (1000 V range)
Voltage display:  ±3% ±3 digits ±0.5% of rated voltage
Max service error:  ±15% ±2 digits.
Short circuit current:  1.5 mA nominal test current
Test current on load:  ≥1 mA at min. pass values of insulation
Output voltage tolerance:  -0% +20% at rated load or less
19.1.2.  Continuity / Resistance
Specfication Detail
Intrinsic accuracy: ±2% ±2 digits (0.01 Ω to 99.9 Ω) 
±5% ±2 digits (100 Ω to 99.9 kΩ)
EN61557 range: 0.1 Ω to 99.9 kΩ
Open circuit voltage  5 V ±1 V
Test current at 200 mA 
(0 Ω to 2 Ω): 
>200 mA @ ≤2 Ω
Test current at 15 mA 
(0 Ω to 2 Ω): 
>15mA @ ≤2 Ω
Max Service error: ±12% ±2 digits.
19.1.3.  Loop test 2Hi (L-PE, L-N, L-L):
Specfication Detail
Intrinsic accuracy:  ±5% ±3 digits
Display range:  0.01 Ω to 1000 Ω
Supply:  48 V to 480 V* (45 Hz to 65 Hz)
Test current high:  4.0 A (at 230 V)
PSCC range:  20 kA
EN61557 range:  0.30 Ω to 1000 Ω
Max service error:  ±10% ±2 digits
* MFT1711 280 V
19.1.4.  Loop test 3Lo, 2Lo (L-E): 
Specfication Detail
Intrinsic accuracy:  0.01 Ω to 39.9 Ω ±5% ±5 digits 
40.0 Ω to 999 Ω ±10% ±5 digits
Display range:  0.01 Ω to 1000 Ω
Supply:  48 V to 280 V (45 Hz to 65 Hz)
Test current:  Pulsed
PFC range:  20 kA
EN61557 range:  0.30 Ω to 1000 Ω
Max service error:  ±10% ±2 digits