10.5  Making a measurement – Three terminal resistance measurement using ART 
1.  Connect the instrument as BELOW. Position the MCC1010 below the point where the Green Lead is attached to 
the Earth Conductor. In practice the MCC1010 can be positioned directly on the Electrode under test, at a point 
below where the green Lead is attached to the the Earth conductor. 
I
CAT IV
300V
12-15V
1.2A
H
SE
Electrode 
under test
Potential 
stake
Current stake
2.  Set the rotary selector switch to the 
3P
   position.
3.  Press and release the 
TEST
 button [by holding the 
TEST
 button, the resistance measurement will be continually 
updated].
The instrument will perform pre-measurement checks, the status of which will be indicated on the display. The three-
terminal resistance reading using ART will be displayed 
Under certain circumstances, the instrument may display a noise warning. This means that interference has been 
detected which may impair the accuracy of the measurement. In particular, the reading could be lower than the 
actual resistance. The resistance of the electrode or system must be verified by an alternative method.
NOTE : 
  The instrument will display the warning triangle and an excessive noise voltage indicator if the ground noise 
voltage is above 20 V pk-pk (7 Vrms). 
  The instrument will display the warning triangle above 2 A – no ART testing is possible under this condition.
  The instrument will display the warning triangle and an over-range condition above 20 A – no ART testing is 
possible under this condition.
  Ensure that the MCC1010 jaw mating surfaces are free of dust and contamination and that they contact 
completely when the MCC1010 is closed.
  Currents carried by conductors in close proximity to the MCC1010 may affect calibration and reduce the 
accuracy of measurements made.
  Re/Rs ratio must be less than 100, where Re = Earth resistance, Rs = Shunt resistance
www.megger.com MFT1800 series
51
Earth resistance measurement