User's Manual
45
Alarm level Condition Message Cause
Alert
Stuck Key
! STUCK KEY
Key pressed for more than 3 minutes.
ESC,OK to clear
Key pressed for more than 3
minutes.
Alert
End of
Suspend
! END OF SUSPEND
Stopped at []. ESC,OK to clear
Pump suspended manually or
automatically.
Alert
Low
Reservoir
! LOW RESERVOIR
[] U or less left [] hours remaining,
replace patch soon. ESC,OK to clear
Low insulin level reached.
Alert
Auto Off
Advisory
! AUTO OFF ADVISORY
Escalates to Auto-Off alarm if not
cleared within 15 minutes. ESC,OK to
clear
Alert indicating that no Patch
status received during the set
time limit.
Alert
Patch
Expiration
Advisory
! PATCH EXPIRATION ADVISORY
Patch expires after [] hours, change
patch soon. ESC,OK to clear
Patch expires in less than the
set time limit.
Alert
Patch
Expiration in
1 Hour
! PATCH EXPIRATION IN 1 HOUR
Patch expires in less than 1 hour,
change patch now. ESC,OK to clear
Patch expires in less than 1
hour.
Alert
Low PDM
! LOW PDM
Low PDM. ESC,OK to clear
The battery is low.
Alert
RF Out of
Range
! RF OUT OF RANGE
Move devices closer and wait 1 minute.
ESC,OK to clear
Pump Base is out of range.