User's Manual
35
3. Pull outward to remove the protective cap from the needle. Save the cap, you will need it later.
Warning: Use with care after removing the needle cap and exposing the fill needle.
4. Determine the amount of insulin you need to insert into the Reservoir Patch. Your healthcare professional will
help you determine the correct amount.
Note: The reservoir requires a minimum of 70 units of insulin to begin operation. The Reservoir Patch can deliver up
to 200 units of insulin.
5. Draw air into the syringe up to the amount of insulin you want.
6. Insert the needle into the insulin vial and push down on the plunger to pressurize the vial.
7. While still holding the plunger rod, flip the vial over so the vial is up, and then slowly pull down on the plunger to
withdraw insulin from the vial into the syringe. Gently tap the side of the syringe to make any air bubbles rise to
the top of the syringe. Slowly push up on the plunger just enough to remove any air bubbles from the syringe.
Slowly pull down on the plunger to fill the syringe to the number of units desired.