User guide
1. Insert a BD™ Test Strip into the test strip slot.
The meter turns on and displays the code
number.
2. Match the code number on the display with
the code on the test strip vial.
If the codes match, you are ready to begin
testing.
If the codes do not match, review “Coding
Your Meter” (refer to page 13).
Getting Started • 17
HOW TO RUN A CONTROL SOLUTION TEST
Meter is off, with no test strip inserted in the test strip slot.
Insert Test Strip
Match Codes
(Example)
16 • Getting Started
CAUTION:
The BD™ Control Solution range printed on the
test strip vial is for control solution only. It is used to test the
performance of the meter and test strip. It is not a recommended
range for your blood glucose level.
Important Control
Solution Information
•Use only the BD™ Control Solution for the test.
• Check the expiration date on the control solution vial.
Do not use the control solution if the expiration date has passed.
•Store only for 3 months after first opening. When you open a new
vial of control solution, count forward 3 months and write that date
on the label of the control solution vial. Discard any remaining
solution after the date you have written on the vial.
•Store the control solution tightly closed at room temperature below
86°F (30°C). Do not refrigerate or freeze.
• Shake control solution well before using.