User guide

IntroductIon
IntroductIon
1312
3
9. While holding the sensor in place,
gently lift the adhesive tab.
Carefully remove the bottom piece
of white paper from under the
adhesive pad. Press the adhesive
against the skin.
10. OPTIONAL Before connecting
the MiniLink transmitter, apply
overtape.
Do not cover the adhesive tab (
1
) or
sensor connectors (
2
) with overtape (
3
).
3
2
Start the sensor
12. While holding the sensor in place,
slide the transmitter onto the
sensor with the other hand by
lining up the connection ends and
gently pressing them until they
click into place.
NOTE! Overtape is not shown in steps 12 - 17.
13. Wait for the green light on the
transmitter to ash green briey.
15. Fold the adhesive tab over the
transmitter.
16. Smooth adhesive onto transmitter.
14. While holding the sensor with one
nger, peel o the paper backing
from the adhesive tab with the
other hand.
Remember. . .
• Ensure that the sensor is fully inserted into the skin. Check that the sensor is not
dislodged when removing the Enlite serter or the introducer needle
• Once inserted, wait at least 5 minutes for the sensor to be wetted with interstitial uid
(it is OK if it takes longer than 15 minutes), before you connect the transmitter
• Use the overtape to reduce sensor pullout
• Proper insertion is important for optimal glucose sensor performance and to prevent
bleeding at the site
Once the sensor is inserted properly under the skin,
• Wait at least 5 minutes; this is the time required for the sensor to be wetted with
interstitial uid
• Remove the MiniLink transmitter from its charger; it will ash green. Wait 1 minute
If the MiniLink transmitter does not ash when connected to the glucose sensor,
• Disconnect the transmitter and place it bacl in the charger, wait until it is fully charged
(i.e. no ashing lights)
• Remove the transmitter from the charger (when fully charged) and wait 1 minute. After
this, connect the transmitter to the sensor and look for the ash
• If you still do not see the ash, your sensor may need more time to wet. With the
transmitter connected to the sensor, drink plenty of uids, wait 2 hours and then
perform the Sensor Start
To start the Medtronic Enlite™ glucose sensor. . .
After you have turned your sensor feature on, enter your MiniLink transmitter ID:
• Go to EDIT SETTINGS in the SENSOR Menu and scroll down to the TRANSMTTER ID
• Use the UP and DOWN buttons to select each digit and press to enter. Be sure to only
enter the numbers
3
Turn ‘Sensor feature on
Paradigm Veo Pump
PRESS:
ACT > Main Menu >
Sensor > Edit Settings >
Sensor: ON > ACT
Paradigm REAL-Time Pump
PRESS:
ACT > Main Menu >
Sensor > Sensor Setup >
Edit Settings >
Sensor: ON > ACT
Guardian Monitor
PRESS:
ACT > Main Menu >
Sensor > Sensor Setup >
Edit Settings >
Sensor: ON > ACT