User guide
82 Chapter 6
Using the Bolus Wizard feature for a Square Wave or Dual Wave bolus
If you are using the Bolus Wizard feature to calculate your Square Wave or Dual Wave bolus
amounts, you will be prompted to enter your BG reading and/or the carb/exchange units you will
eat. The Bolus Wizard feature will use this input to calculate your suggested correction and/or food
bolus amount. If you do not want to use the Bolus Wizard feature estimate, you can change it, if
desired.
1 The Bolus Wizard feature must be turned on and the settings must be programmed. Also, make sure
the dual/square option is turned on.
NOTE - If you want to use the pump-to-meter link, make sure the meter option is on. Refer to
the section “Meter option” for instructions.
2 After you enter your BG and/or food entry, review the
information in the ESTIMATE DETAILS screen. Press ACT to
continue to step 3. If you need to make any changes, press
ESC to return to the ENTER BG screen and make changes as
necessary.
3 In the next screen, select Square Wave Bolus
or Dual Wave Bolus as desired and press ACT.
NOTE - If you are using the Bolus Wizard
feature and it calculates that your bolus includes a portion to
correct your high BG, the Square Wave bolus will not be
available. This helps you to select a bolus type (Normal or Dual
Wave) that has an immediate delivery option to cover your high
BG.
(values shown are
for example only)
ESTIMATE DETAILS
Est total: 4.0U
Food intake: 15gr
(Meter) BG: 227
Food:3.0U
Correction:2.5U
Active Ins: 1.5U
ACT to proceed,
ESC to back up
(using the Bolus Wizard
feature)
(without the Bolus
Wizard feature)
BOLUS EST: 4.0U
Normal Bolus
Square Wave Bolus
Dual Wave Bolus
BOLUS TYPE
Normal Bolus
Square Wave Bolus
Dual Wave Bolus
ACT
ACT
BOLUS EST: 4.0U
Normal Bolus
Dual Wave Bolus