User guide
78 Chapter 6
Dual Wave/Square Wave bolus On/Off
NOTE - It is important that you consult with your healthcare professional before using a Square
Wave or Dual Wave bolus. You should be familiar with the basic functions of your pump before
exploring these options.
To set up a Dual Wave or Square Wave bolus, you must first turn on the Dual/Square bolus option. If
the option is off, a Dual Wave or Square Wave bolus cannot be programmed or delivered.
1 Go to the DUAL/SQUARE OPTION screen.
Main > Bolus > Dual/Square Bolus
2 Select On and press ACT. The feature is now on. Exit the menus.
Square Wave or Dual Wave bolus without Bolus Wizard feature
NOTE - To deliver a Square Wave or Dual Wave bolus, the Dual/Square bolus option must be on.
1 Make sure the Dual/Square option is on.
2 Calculate your food and/or correction bolus amount.
3 Press on your pump, or go to the BOLUS MENU and select Set Bolus (or Manual Bolus), then
press ACT.
4 The BOLUS TYPE (or MANUAL BOLUS TYPE) screen will appear.
DUAL/SQUARE OPTION
Off
On