User guide
Using the Bolus Wizard feature 69
Meter rules
If you want your pump to communicate with the meter, the following conditions must apply:
1 The meter option must be turned on and programmed. Refer to the instructions in this section.
2 Your pump must be within 4 feet (1.2 meters) of your meter to receive the BG reading.
3 The pump cannot have a Low Battery alert condition.
4 When programming a bolus, the BG measurement from the meter will appear as the default BG
value on the ENTER BG screen. The pump will not display a reading that is older than 12 minutes on
the ENTER BG screen.
5 Make sure the pump’s Meter Option is set to Off while onboard aircraft.
6 Do not use the RF meter to send your BG readings to the pump while onboard aircraft. Manually
enter your BG.
Add, delete, review meter IDs
The meter programming screens are very similar to those for the remote control. Make sure to
select Meter Options (in the UTILITIES MENU) when programming your meter.
If you are not sure that your meter ID is entered in your pump, check the REVIEW METER ID screen.
You have to turn on the meter option to add, delete or review the meter ID(s) programmed in your
pump.
Main > Utilities > Meter Options
1 In the METER OPTION screen, select On and
press ACT. The METER ID MENU will appear.
CAUTION: The pump will not receive signals from the linked meter while it has a Low Battery
alert condition. To ensure the meter communicates with the pump, make sure the
pump does not have a low battery. (Replacing the low battery with a new battery will
restore meter-pump communication.)
METER OPTION
Off
On
ACT
METER ID MENU
Add ID
Delete ID
Review ID