Using the Bolus Wizard feature 63
6 The screen will return to the EDIT SETTINGS
screen. Set your BG Units as described in the
next section.
NOTE - Carb ratio values are normally between 5–50 grams/u or 0.3–
3.0 u/exch. If your ratio value is outside the range, this warning
message will appear on the screen. Press ESC to correct or ACT to
continue.
To set just one ratio: To set more than one carb ratio:
a. Press ESC. a. Set the start time
for the next ratio
and press ACT.
b. Set the value for
the next ratio and
press ACT.
c. Repeat steps a
and b to set more
ratios.
d. Press ESC when
you are done.
WARNING
A carb ratio of XX is valid
but outside the usual
range of 5-50.*
ESC to correct,
ACT to continue
(* 0.3 - 3.0 for exch)
_ : _
__
(flashing)
using grams
(flashing)
using exchanges
SET START TIME 2
grams
/U
_ : _
__
SET START TIME 2
U/exch
(flashing)
set then press ACT
using grams using exchanges
* Depending on how many ratios you set,
this number can be 2 through 8.
_ : _
__
SET START TIME 2
grams
/U
_ : _
__
SET START TIME 2
U/exch
7:00A
__
SET CARB RATIO 2*
grams
/U
7:00A
__
SET EXCH RATIO 2*
U/exch
- >
EDIT SETTINGS
Wizard:
On
Carb Units:
grams
Carb Ratios: