User guide

Using the Bolus Wizard feature 59
More about the Bolus Wizard feature
About high or low BG levels
If your BG is less than 70 mg/dL (3.9 mmol/L) or
more than 250 mg/dL (13.9 mmol/L), the Bolus
Wizard feature screen will notify you and give
instructions. Read the instructions and press ACT or
ESC to clear the message. You can continue
programming and deliver your bolus.
About maximum delivery
Bolus Wizard feature will not deliver more than the limit set for
your maximum bolus. If the Bolus Wizard feature calculates a bolus
amount that is larger than your max bolus limit setting, the
message, “MAX BOLUS EXCEEDED” will appear. If this happens, do
these steps:
1 In the MAX BOLUS EXCEEDED screen, press ACT to continue your
bolus programming. The estimate and maximum bolus amounts
will appear for your information. Continue to step 2.
If you do not want to continue, press ESC to cancel and the screen
will return to the ENTER BG screen.
2 In the EST: MAX screen, press ACT again to continue your bolus
programming. If desired, press ESC to cancel and the screen will
return to the ENTER BG screen.
NOTE - Your pump will only deliver up to your maximum bolus limit setting. For example: The
Bolus Wizard feature estimate is 30 units and your max bolus limit is 25 units. When you press
ACT, your pump will only deliver 25 units, and will notify you that your bolus estimate exceeds
your max bolus.
LOW BG
DO NOT BOLUS UNTIL BG
IS NORMAL.
Treat low BG.
Monitor BG.
HIGH BG
Check for occlusion
Check ketones.
Consider insulin
injection.
Monitor BG.
ACT or ESC
- - -
MAX BOLUS
EXCEEDED
ACT
ESC
to cancel
EST: 30.0U
MAX: 25.0U
ENTER BG
mg/dL*
*or mmol/L