User guide

Starting on insulin 45
Rewinding your pump
Before you continue, make sure the pump is NOT connected to your body.
1 If you removed your reservoir and are replacing it, go to the REWIND screen.
Main > Prime > Rewind
2 In the REWIND screen, press ACT to start the rewind
process. The REWINDING screen will appear while
the pump rewinds.
3 After the pump rewinds the PREPARING TO PRIME
screen will appear.
NOTE - If you press ESC
,
the PRIMING
STOPPED screen displays. This screen guides
you back to the priming screens. Once you start
a rewind, you cannot cancel it.
If you are practicing, do these steps:
a. Do NOT insert the reservoir in your pump. Make sure the shipping cap is installed in the
reservoir compartment.
b. Continue with the manual prime instructions described in the “Manual prime” section on
page 47.
If you are not practicing, continue to the next section, “Inserting the reservoir in your pump.”
WARNING: Make sure the infusion set is disconnected from your body before you rewind or prime
the pump. Never insert the reservoir into the pump while the tubing is connected to
your body. Doing so could result in an accidental infusion of insulin.
Disconnect from Body
Press ACT to Rewind
Wait for notification
REWIND
REWINDING
ACT
PRIMING STOPPED
Rewind Pump
Status Screen
PREPARING TO PRIME
Disconnect from Body
Insert/Lock Reservoir
Hold ACT to Prime
ESC
ACT
Continue Priming