User guide

44 Chapter 4
Changing your infusion set
Removing the reservoir
Each time you remove and replace a reservoir in your pump, you have to rewind and prime your
pump. Priming requires insulin.
11 Tap the side of the
reservoir to remove any air
bubbles.
12 To purge air bubbles that
have risen to the top of the
reservoir, push up on the
plunger until you see insulin
in the tubing.
13 Without pulling, turn the
plunger rod counter-clockwise
to remove it from the
reservoir.
1 Remove the entire infusion set from your body.
2 If attached, remove the activity guard.
3 Turn the tubing connector half-turn counter
clockwise, then pull the reservoir and connector
out from the pump.
4 Safely dispose of the used reservoir and infusion set items in a sharp’s container.
5 You must now rewind your pump as described in the next section.
tubing connetor
half-turn counter-clockwise