User guide

Starting on insulin 43
5 While still holding the
plunger rod, flip the vial
over so the vial is up,
slowly pull down on the
plunger to fill the
reservoir.
6 Gently tap the side of the
reservoir to make any air
bubbles rise to the top of the
reservoir.
7 Slowly push up on the plunger
just enough to remove any air
bubbles from the reservoir.
8 Slowly pull down on the
plunger to fill the reservoir to
the number of units desired.
9 With the vial down, hold the transfer guard.
Turn the reservoir counter-clockwise then
pull straight up to remove it from the
transfer guard.
10 Place the tubing connector onto the reservoir.
Turn the connector clockwise pressing gently
against the reservoir until you feel it slide in.
Push in and continue turning until the reservoir
and the connector lock with a click.
“CLICK”