User guide
Pump specifications 143
Bolus Wizard feature examples
Settings:
➠ Carb ratio: 30 grams/unit ➠ Insulin Sensitivity: 40 mg/dL/unit
➠ BG Target: 90-120 mg/dL ➠ Active Insulin Time: 6 Hours
#1: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and does not enter BG.
#2: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 200.
#3: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 70.
#4: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 100.
60g
30g/u
(food estimate)
= 2 units
= 2 unitsestimate
60g
30g/u
200mg/dL - 120mg/dL
40mg/dL/u
= 2 + 2
= 4 units
estimate
+
(food estimate)
= 2 units
= 2 units
(correction)
= 4 units
60g
30g/u
70mg/dL - 90mg/dL
40mg/dL/u
-20mg/dL
40mg/dL
+
(food estimate)
= 2 units
= 2 + (-0.5)
= 1.5 unit
= 1.5 unit
(correction)
= -0.5 unit=
estimate
60g
30g/u
= 2 units
+
Correction is 0 because the current BG
reading is between the BG High and Low
(correction)
(food estimate)
= 2 + 0
= 2 unit
= 2 unitestimate