User guide

Pump specifications 137
Easy bolus
Features programming using audible tones (or vibrate pulses) in user-determined increments. Beep mode range: 0
to maximum bolus; Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first.
Default step increment: 0.1 unit
Step size < maximum bolus.
Adjustable step size: 0.1 to 2.0 units per step.
Accessible from the remote control or pump buttons.
Infusion pressure
Maximum infusion pressure and occlusion pressure: 13.7 PSI (94,46 kPa).
Insulin sensitivity
Maximum settings: 8
Factory default: 50 mg/dL (2.8 mmol/L)
Low resv (reservoir) warning
Values are based on displayed amount, not actual amount.
Meter value
The BG measurement received from the linked meter. Appears in the ENTER BG screen during bolus programming.
Appears on the screen when the pump is idle at the HOME screen.
Expiration: 12 minutes;
Range: 20 – 600 mg/dL (1.1 - 33.3mmol/L)
Maximum meter ID entries: 3
Normal bolus
Range 0.1 – 25.0 units of insulin (limited by maximum bolus setting).
Range:10 – 400 mg/dL
(0.5 – 13.9 mmol/L)
Warning limits:less than 20 or greater than 100 mg/dL
(less than 1.1 or greater than 5.6 mmol/L)
Time: 2 – 24-hours, and
@ 1:00 hour before empty
08:00 hours (default when time is
selected)
Units: 5 – 50 units, and
@ 1/2 amount remaining
20 units (factory default)