User guide
136 Chapter 11
Bolus units
Increments: 0.1 units
Bolus Wizard feature
(see end of this section for information)
Carb ratios
Carb units
The food entry when using Bolus Wizard feature
grams: 0 – 300 (increments: 1 gram)
exchanges: 0.0 – 20 (increments: 0.5 exch)
Daily totals
Maximum records shown: 31 days of data, maximum display: 999.95 units/day. Accuracy: +0/–0.05 units
Default screen
The HOME screen. For most screens, if no buttons are pressed for 30 seconds, the pump returns to this screen.
Pump motor
The pump motor has a unique, patented design with an integrated safety check system. The system delivers in
precise increments.
Dual Wave bolus
Delivers a Normal bolus followed by a Square Wave bolus (limited by maximum bolus).
Maximum ratio settings: range: warning limits:
8
3 – 150 grams/u
0.1 – 5.0 u/exch
less than 5 or greater than 50 grams/u
less than 0.3 or greater than 3.0 u/exch