User guide

Pump specifications 135
Chapter 11
Pump specifications
This section provides detailed information on specifications related to your pump. The safety features of your
pump are individually listed and described.
Alarms and error messages
Indicators: audible tone (beep) or vibration (silent)
All alarms and errors show messages on the pump’s screen and provide instructions on what to do. Unresolved
alarms will escalate to siren mode for added safety.
Alarm history
Maximum records shown: 36
Backlight
LCD (Liquid Crystal Display) type
Time-out: 30 seconds
Basal
Delivery: 0.05 – 35 units/hour (maximum units: 35/hour)
Factory maximum setting: 2.0 units/hour
Increments: 0.05 units
3 patterns maximum, each with 48 rates
BG target
Maximum targets: 8
Bolus delivery
Bolus history
Maximum records shown: 24
range:60 – 250 mg/dL
(3.3 – 13.9 mmol/L)
warning limits:less than 90 or greater than 140 mg/dL
(less than 5.0 or greater than 7.8 mmol/L)
Insulin delivered/stroke Fluid delivered/stroke Time between/delivery strokes Delivery rate (per minute)
0.05u 0.5 µL 2 seconds 1.5u