User guide
Troubleshooting 131
Max delivery
This alarm alerts you when you have taken more insulin than expected based
on maximum bolus and maximum basal rates.
Motor error
Insulin delivery has stopped. This alarm occurs if your pump detects a motor
error.
No delivery
Insulin delivery has stopped. This alarm occurs if your pump detects a
blockage.
No reservoir
The reservoir is not inserted correctly or no reservoir has been inserted.
Off no power
The battery is dead. Replace battery immediately. Follow the directions on
the screen. Check to make sure that the time is correct on the screen. Reset
the time if necessary.
MAX
DELIVERY
Exceeded 1 hour
max delivery
Check BG
3:25P
ESC, ACT to clear
MOTOR
ERROR
Delivery stopped
Disconnect set
ESC, ACT to clear
3:25P
Delivery stopped
Change entire set
Check BG
See user guide
to troubleshoot
ESC, ACT to clear
NO
DELIVERY
11:17A
NO
RESERVOIR
Delivery Stopped
Change reservoir set
ESC, ACT to clear
4:36P
0% battery life
Delivery stopped
Replace battery now
04:42
OFF
NO POWER
Use 1 AAA
ESC, ACT to clear