130 Chapter 9
Check settings
When this alarm is active, you should check and/or reprogram your pump
settings, including the time/date.
E (Error)
After receiving this alarm, note the error number then call the Medtronic
MiniMed 24 Hour HelpLine for assistance. An error alarm will show an āEā
followed by two numbers. E-alarms cause all insulin delivery to stop, the
pump resets, and all your settings are cleared.
Empty reservoir
There is no insulin in the reservoir. Change the reservoir immediately.
Failed batt test
The pump tests the voltage of each battery installed. This test ensures a
battery with low voltage is not used. If the battery does not have enough
voltage, this alarm will occur. The pump will not function and the battery
must be replaced. (Always make sure that you install a NEW battery into the
pump.)
Is priming complete?
If you manually prime your pump with more than 30u insulin, this message
appears. Press ESC, ACT to clear the message. If manual prime is complete,
press ESC. If the manual prime is not complete, press and hold ACT until
manual prime is complete.
8:35A
CHECK
SETTINGS
Delivery stopped
Reprogram settings
ESC, ACT to clear
Exx
XX indicates the error
number
5:30A
Settings cleared
Reprogram settings
Call help-line
for assistance
ESC, ACT to clear
3:36P
EMPTY
RESERVOIR
Delivery stopped
Change reservoir
ESC, ACT to clear
8:35A
FAILED
BATT TEST
Delivery stopped
Replace battery now
Use 1 AAA
ESC, ACT to clear
IS PRIMING
COMPLETE?
Disconnect from body
See user guide
for instruction
ESC, ACT to clear