116 Chapter 8
Selftest
Selftest is a safety utility that allows you to check if your pump is operating properly. This self-
diagnostic feature can be used for maintenance or to check your pump if it operates unusually.
During selftest, your pump will automatically run internal tests, including a check for proper
operation of the beep and vibrate modes. The selftest is additional to the routine tests that run
independently while the pump operates.
Contact the Medtronic MiniMed 24 Hour HelpLine if any of the tests do not occur as described here.
NOTE - If the pump detects a condition such as low battery, the selftest will not finish. A message
will appear to show the condition that caused the test to stop.
1 Go to the UTILITIES MENU. Select Selftest and
press ACT.
Main > Utilities > Selftest
2 As part of the selftests, the pump will do these
tests:
NOTE - Periodically, you will hear beeps as different mechanisms in the pump are being tested.
3 After the selftest is finished, TEST COMPLETE will
appear on the screen. The screen will return to the
UTILITIES MENU, then to the HOME screen.
a. Screen Test:
The screen will appear
all black as shown
here.
b. Selftest:
The pump will count
down from 10.
c. Tone Test:
You should hear beeps.
d. Vibrate Test:
You will feel
vibrations.
Basal
Prime
ACT
UTILITIES MENU
Remote Options
Block
MAIN MENU
Utilities
Selftest
10
SELFTEST
3
TONE TEST
VIBRATE TEST
1