User guide
Optimizing pump therapy 85
Step value setup
You can set the step value from 0.1 to 2.0 units (factory setting: 0.1). Set the step to a number that
is convenient to use and easy to multiply.
2 In the EASY BOLUS ENTRY screen, the “step”
value will appear flashing. Change the value
and press ACT (The step value is the
increment you will use for your Easy bolus.)
3 The screen will return to the BOLUS MENU.
Your step amount is now programmed and
Easy Bolus is ready to use. Exit the menus.
Deliver Easy Bolus
Practice using the Easy Bolus feature while looking at the pump screen as you count the beeps.
After you are familiar with Easy Bolus, you can use the audible tones for bolus delivery without
having to look at the screen.
NOTE - Easy bolus only works from the HOME screen.
1 From the HOME screen, press . The SET EASY BOLUS
screen will appear. The single “step” value will be flashing.
4
3
2
1
0 1.00.5 1.5 2
Example:
2.0 unit Easy bolus with a step
amount set at 0.5 units
0
Total # Steps = 4
Total # Button Presses = 4
Total # Units
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
/
(flashing)
EASY BOLUS ENTRY
0.
1 U
ACT
ESC
BOLUS MENU
Easy Bolus
BG Reminder
EASY BOLUS ENTRY
Dual/Square Bolus
0.
5 U
(flashing)
SET EASY BOLUS
U
0.1