Operating instructions
LIFEPAK CR–T Trainer Operating Instructions 25
©2002-2007 Medtronic, Inc.
Caring for the LIFEPAK CR–T AED Trainer
Caring for the LIFEPAK CR–T AED Trainer
Table 13 Setup Options Troubleshooting
Observation Possible Cause Corrective Action
Voice prompts are out of sequence • Incorrect operating
setting options
selected
• Review setup procedures in
Section 3, beginning on page 15.
Voice prompts are incomplete • Trainer damaged • Contact your local Medtronic
representative.
CONNECT ELECTRODES sequence
repeats after training electrodes
are connected to the Trainer
• Instructor did not press
remote control to begin
scenario sequence
• Damaged remote
control
• Infrared sensor not
working properly
• Press the ELECTRODE PADS button
once on the remote control to
initiate voice prompts.
• Contact your local Medtronic
representative.
• Clean or clear the signal sensor, if
obstructed.
• Install remote control cable.
• Contact your local Medtronic
representative.
Table 14 Remote Control Troubleshooting
Observation Possible Cause Corrective Action
Trainer does not respond to remote
control button presses
• Infrared sensor is not
receiving remote
control signal
• Reposition remote control relative
to the Trainer.
• Remove anything blocking the
signal sensor.
• Install remote control cable.
• D-cell batteries
depleted or damaged
• Replace D-cell batteries.
• Remote control
damaged
• Replace remote control.
• Remote control
batteries low
• Replace remote control batteries.
Table 15 Training Electrodes Troubleshooting
Observation Possible Cause Corrective Action
Training electrode pads do not
adhere to manikin
• Adhesive
contaminated
• Replace training electrode
packet.
Training electrode connector will
not plug into the Trainer
• Training electrode
connector damaged
• Trainer electrode
receptacle damaged
• Replace training electrode
packet.
• Contact your local Medtronic
representative.