Operating instructions

LIFEPAK CR–T Trainer Operating Instructions 11
©2002-2007 Medtronic, Inc.
Using the LIFEPAK CR–T AED Trainer
Using the LIFEPAK CR–T AED Trainer
T1
Change shock to no
shock advised
The NO SHOCK ADVISED button can be used only during
simulated analysis.
Press the
NO SHOCK ADVISED button to change any shock within
a shock protocol to
NO SHOCK ADVISED.
F3
Simulates the
CHARGE-PAK needs to
be replaced
The
CHARGE-PAK button can be used only when the Trainer is in
simulated OFF status (
MASTER POWER switch ON, LID
RELEASE/ON-OFF
button OFF).
Press the
CHARGE-PAK button once to activate the CHARGE-PAK
indicator in the readiness display.
The
CHARGE-PAK indicator automatically appears on the
readiness display after simulating a patient use.
To clear the
CHARGE-PAK indicator from the readiness display,
remove and replace the CHARGE-PAK. The instructor can also
clear the indicator by pressing the button a second time.
F4
Simulates a call for
service situation
The
WRENCH button can be used only when the Trainer is in
simulated OFF status.
Press the
WRENCH button once to activate the wrench indicator
in the readiness display.
To clear the
WRENCH indicator from the readiness display, press
the button a second time.
T2
Attention
The
ATTENTION button can be used when the Trainer is ON and
when it is in simulated OFF status.
Press the
ATTENTION button once to activate the ATTENTION
indicator in the readiness display. This simulates a low internal
battery condition (simulating a situation in which the actual
defibrillator would have a minimum of 6 shocks or 42 minutes of
monitoring time available for use).
To clear the
ATTENTION indicator from the readiness display,
press the
ATTENTION button a second time.
C1
Start or stop simulated
motion
The
MOTION button is used only during simulated analysis.
Press the
MOTION button to start simulated motion. The Trainer
interrupts simulated analysis and delivers
MOTION DETECTED
prompts for 10 seconds or until the instructor presses the
MOTION button again.
C2
Pause
The
PAUSE button suspends the Trainer in its present state and
causes the LED display to illuminate red until operation is
resumed. This allows the instructor to interject other instruction,
if necessary. The Trainer will pause for approximately
30 minutes before it turns completely off.
Pressing the
PAUSE button a second time resumes Trainer
operation.
Table 6 Remote Control Button Functions (Continued)
Button Function Operation and Description