User's Manual
Table Of Contents
- Table of Contents
- Instructions for use
- Getting started
- Using the SEEQ MCT System during the monitoring period
- What to do at the end of monitoring
- Indications for use, contraindications, and precautions
- Services for physicians
- Specifications, compliance, and symbols
- Frequently asked questions
- What is a Wearable Sensor?
- How long will my Wearable Sensor last?
- What information is transmitted to my physician?
- How can I report my symptoms?
- Can I take a shower while wearing my Wearable Sensor?
- Will I need to change the battery in the Transmitter or the Wearable Sensor?
- How close must I be to the Transmitter to ensure that the data collected by my Wearable Sensor is transmitted?
- How often do I need to change the Wearable Sensor device?
- What if the Wearable Sensor causes my skin to itch?
- Should I carry the Transmitter with me when I travel?
- Will I need to notify security screeners about my Wearable Sensor?
- Can I wear a Wearable Sensor through an electronic antitheft system, such as in a store?
- Can I use microwave ovens or TV remotes while wearing a Wearable Sensor?
- Can I be close to wireless phones, WiFi, or other electromagnetic devices?
- Can I carry my cellphone while wearing the Wearable Sensor?
- Will hot or cold environments affect the Wearable Sensor performance?
- Is my medical data protected during transmission?
SEEQ MCT Instructions for use English 17
M958549A001 Rev B
Refer to doc # A73903 for Printing
Instructions.
M958549A_ch.fm 3/12/15 2:34 pm
4" x 8" (101 mm x 203 mm)
Medtronic Confidential
mdvtldct_R1.5
Specifications, compliance, and symbols
Specifications
The following performance specifications are at 68 °F (20 °C) unless
otherwise stated.
Wearable Sensor Transmitter
Shelf life Refer to Wearable
Sensor pouch label
N/A
Battery Charger power
requirement
N/A 100-240 VAC,
50/60 Hz
Battery life 7.5 days
(180 hours),
non-rechargeable
Provides 12 hours of
function before
recharging
Operating temperature 32 °F to 113 °F
(0 °C to 45 °C)
32 °F to 113 °F
(0 °C to 45 °C)
Maximum temperature of
the applied part
111.2 °F (44 °C) N/A
Storage temperature
(power off)
-4 °F to 149 °F
(-20 °C to 65 °C)
-4 °F to 149 °F
(-20 °C to 65 °C)
Operating humidity 10% to 95% 10% to 95%
Storage humidity 5% to 95% 5% to 95%
Operating atmospheric
pressure
700 hPa to
1060 hPa
700 hPa to
1060 hPa
ECG
Sampling rate
Digital resolution
Input dynamic range
Input offset dynamic range
200 Hz (+/-5%)
16bits
+/-5 mV
+/-300 mV
N/A
Impedance measurements
Peak current injection
RMS current injection
40 uA
29 uA
N/A
Measurement ranges
Heart Rate 25 to 250 bpm N/A
Data storage
Capacity 7.5 days N/A
Weight 1.8 oz / 50 g max 5.3 oz / 150 g max
Communication means Bluetooth between
Wearable Sensor
and Transmitter
Cellular Phone between
Transmitter and Server