Users Manual
Table Of Contents
Note: The Auto Suspend feature continues working when your pump
switches to SmartGuard Auto Mode.
To set up Auto Suspend:
1. Press
and go to the Auto Suspend screen.
Options > Delivery Settings > Auto Suspend
2. Select Alarm.
3. Select Time and enter the number of hours you want to set.
4. Select Save.
Block Mode
The Block Mode feature lets caregivers, such as parents of a young child, restrict
access to critical pump settings.
WARNING: Always monitor the pump when it is used in Block
Mode. You can manually suspend while in Block Mode. This could
result in hyperglycemia and ketoacidosis.
When Block Mode is on, you cannot start a new bolus delivery, start a new basal
pattern, or start a new temp basal delivery. Any previous bolus and basal deliveries
continue normally, and the pump user can stop a bolus delivery at any time.
When your pump is in Block Mode, you can suspend insulin delivery, receive
sensor glucose (SG) values, receive blood glucose (BG) values from your Accu-Chek
Guide Link meter, review history, test the pump, and clear alarms and alerts.
However, you cannot change any settings.
Note: The Block Mode feature has some differences when your pump is
in Auto Mode. See Block Mode when in SmartGuard Auto Mode, on
page 228.
To turn Block Mode on or off:
1. Press
and go to the Block Mode screen.
160 Chapter 9