Event Markers
The Event Markers feature lets you electronically save certain types of information.
When using this feature, enter events when they happen because the system
records the time of the entry. You cannot edit entries after you enter the
information into your pump. You can view your saved events in the Daily History
screen.
The information you entered can be sent to CareLink Personal software, where it
can be used to generate reports you can share with your healthcare professional.
To enter Event Markers:
1. Press
and go to the Event Markers screen.
Options > Event Markers
2. Select and enter event information for any of the following categories:
BG If you are not recording your BG meter readings in your
pump by entering them manually or by using the Bolus
Wizard feature, Auto Mode Bolus, or an Accu-Chek Guide
Link meter, you can enter them in this screen. If you use a
sensor, you may use a BG meter reading you enter in this
screen for calibration. You can also enter non-calibration
BG meter readings, such as those readings taken when
eating or when your BG is rising or falling rapidly.
Injection
Enter the number of units of any insulin you have given
by injection.
Note: Insulin units entered using the
injection event marker are not added to
your Active Insulin amount tracked on the
pump.
146 Chapter 7