WARNING: Never leave your pump on the Fill Cannula? screen.
Insulin delivery is suspended while on the Fill Cannula? screen.
Always finish filling your cannula or return to the Home screen to
avoid continued insulin delivery suspension. Failing to do this can
result in hyperglycemia.
To fill the cannula:
1. After you fill your tubing and insert your infusion set, the Fill Cannula? screen
appears.
Note: If your screen turns off before you are ready to fill your
cannula, press any button on your pump to turn it on again.
2. To fill your cannula now, select Fill. If you use a steel needle infusion set,
there is no cannula to fill. Select Done.
The Fill Cannula screen appears.
3. Adjust the Fill amount for your particular infusion set, and then select Fill
Now. If you are unsure about the fill amount, see the instructions that came
with your infusion set.
■
Reservoir and infusion set