Users Manual
Table Of Contents
Note: If you press the Back button after the loading process begins,
a Loading incomplete alarm will occur.
When the loading process is completed, the following screen appears.
7. Select Next to continue.
8. Follow the instructions in the next section to fill the tubing with insulin.
Filling the tubing
You need to fill the infusion set tubing with insulin before you insert the set into
the body.
WARNING: Always make sure the infusion set is disconnected from
your body before you rewind your pump or fill the infusion set
tubing. Never insert the reservoir into the pump while the tubing is
connected to your body. Doing so could result in an accidental
infusion of insulin, which can cause low blood glucose.
■
Reservoir and infusion set
Reservoir and infusion set 125