When you use the Bolus Wizard feature, the Bolus Wizard calculator uses your
current active insulin value to determine if there is an active insulin adjustment
needed. The active insulin adjustment calculation considers both the bolus insulin
that has previously been delivered (the amount shown on the Home screen), as
well as any insulin that will be delivered by an active Square Wave bolus.
WARNING: Do not use the Bolus Wizard feature to calculate a bolus
for a period of time after giving a manual injection of insulin by
syringe or pen. Manual injections are not accounted for in the
active insulin amount. Therefore, the Bolus Wizard feature could
prompt you to deliver more insulin than needed. Too much insulin
can cause hypoglycemia. Consult with your healthcare professional
for how long you need to wait after a manual injection of insulin
before you can rely on the active insulin calculation of the Bolus
Wizard feature.
Bolus Wizard feature alerts
When you use the Bolus Wizard feature, there may be times when you see one of
the following:
Alert: What it means: What to do:
High BG Your BG meter reading is above
250 mg/dL.
• Check infusion set.
• Check ketones.
• Consider an insulin injection.
• Monitor your BG.
Low BG Your BG meter reading is below
70 mg/dL.
Treat your low BG. Do not give
yourself a bolus until your BG
returns to normal.
■
Bolus
Bolus 97