Product Info

Test Description
Tone An alert tone, an Easy Bolus step tone, and an alarm tone are
generated.
The pump performs a series of tests as listed in the previous table. Self Test
requires you to observe the pump during the test.
To run the Self Test:
1. Press
and go to the Self Test screen.
Options > Utilities > Self Test
A message indicates that the Self Test is in progress.
Self Test takes up to two minutes to complete. During that time, the display
briefly turns white, the notification light blinks, the pump vibrates, and the
pump beeps.
2. If Self Test does not detect a problem, the display returns to the Utilities
screen.
If Self Test detects a problem, a message appears with more information
about the problem. If Self Test displays an error message or you observe the
pump not behaving as indicated during the test, contact 24-Hour Technical
Support.
Sensor Demo
Sensor Demo lets you see what the Home screen would look like if you were
using the optional CGM feature. For more information about sensor graphs, see
The sensor graph, on page 211.
WARNING: Do not use Sensor Demo to make any decisions related
to your therapy. Information seen in the Sensor Demo is not real
data. It is an example of the type of information you can access
when using the sensor feature. Making treatment decisions based
on data that is not real can cause hypoglycemia or hyperglycemia.
To view the sensor graph example screens:
1. Press
and go to the Sensor Demo screen.
General settings
General settings 167