Upload to CareLink software
Upload system data to CareLink software with the MiniMed Mobile app or the blue
adapter (ACC-190).
The following steps are instructions to upload system data to CareLink software
with the blue adapter. Refer to the MiniMed Mobile app user guide for instructions
to upload system data to CareLink software with the app.
To upload to CareLink software with the blue adapter:
1. Press and hold
, or press and go to the CareLink screen.
Options > Utilities > CareLink
2. Follow the instructions on the CareLink uploader.
3. The CareLink uploader tells you to enter a pump code if the pump is new to
the CareLink account. Enter the Pump Code from the CareLink screen on the
pump.
4. Select Next on the CareLink uploader.
5. Select Upload Now on the pump screen.
Self Test
Self Test is a safety utility that lets you check if your pump is operating properly.
This self-diagnostic feature can be used for maintenance or to check that your
pump is operating properly. Self Test is additional to the routine tests that run
independently while the pump operates.
Note: Your insulin delivery suspends for up to two minutes while your
pump runs a self test.
Self Test includes the following tests:
Test Description
Display The display turns on for up to 45 seconds.
Notification light The notification light turns on for three seconds, and then turns it
off.
Vibration Two vibration tones are generated.