Product Info

"Warm up..." appears on the Home screen until the sensor is ready for first
calibration.
Reconnecting the sensor
There are times when you remove the transmitter from an inserted sensor. After
you reconnect the transmitter to the sensor, the pump detects the connected
transmitter. A "Sensor connected" message appears.
To reconnect a sensor:
1. Select Reconnect Sensor.
The "Sensor warm-up started" message appears.
Note: It may take up to five minutes for the "Sensor warm-up
started" message to appear.
2. Select OK.
"Warm up..." appears on the Home screen until the sensor is ready for its first
calibration.
Calibrating your sensor
Calibration is the process of entering a BG meter reading to calculate SG values.
You must calibrate your sensor regularly to ensure you continue to receive SG
data. For details, see Guidelines for calibrating, on page 207.
Within two hours after you use your pump to start the sensor, your pump displays
a Calibrate now alert to let you know that a calibration is due. This BG meter
reading is the first calibration for your sensor. It takes up to five minutes after
calibration to see the first SG reading on your Home screen. You enter your
second calibration within six hours after your first calibration.
After you have entered your first two calibrations, you must calibrate your sensor
again within 12 hours. If you do not enter a BG meter reading within 12 hours,
your pump displays the Calibrate now alert and stops calculating SG values until a
calibration BG is successfully entered. The sensor must be calibrated at a minimum