Note: When you enter your settings, you first define the time segment,
and then select the low SG settings you want during that time segment.
To set up the low SG settings:
1. Press
and go to the Low Setup screen.
Options > SmartGuard > Low Setup
The Low Setup screen appears.
2. Select the time segment. The End time flashes.
The Start time of the first time segment is always 12:00 A. You can set up to
eight time segments, each with a different low limit. If you set more than one
time segment, the time segments must cover a 24-hour period.
3. Set the End time.
4. Set your low limit. Enter a value in increments of 5 mg/dL from 50 to 90
mg/dL.
5. Select the arrow to the right of the End time to select the low SG settings for
this time segment.
A screen appears and shows the available settings for the selected time
period.
6. Set the following as desired:
196 Chapter 10