Product Info
Table Of Contents
The High Setup screen appears.
2. Select the time segment. The End time starts flashing.
The Start time of the first time segment is always 12:00 A. You can set up to
eight time segments, each with a different high limit. If you set more than
one time segment, the time segments must cover a 24-hour period.
3. Set the End time.
4. Set your High limit. You can enter a value from 100 to 400 mg/dL, in
increments of 5 mg/dL.
5. Select the arrow to the right of the End time to select the high alerts for this
time segment.
A screen appears and shows the high alerts for the selected time segment.
6. Set the following alerts as desired:
a. Select Alert before high to receive an alert before you reach your high
limit.
b. Set the Time before high option between 5 to 30 minutes to receive an
alert before you reach your high limit.
c. Select Alert on high to receive an alert when you reach your high limit.
d. Select Rise Alert to receive an alert when your SG is rising quickly.
Go to step 11 if you do not select Rise Alert.
■
Setting up CGM
Setting up CGM 193