Note: The maximum amount of time the Suspend on low feature will be
unavailable is four hours. After this time period, the Suspend on low
feature automatically enables.
When the SmartGuard suspend features are unavailable, the SmartGuard suspend
icon on the Home screen appears gray
.
When a Suspend on low event occurs and you respond within two hours and:
• Stay suspended for the two-hour maximum suspend time, the SmartGuard
suspend features will be unavailable for 30 minutes after your basal insulin
delivery resumes.
• Your insulin automatically resumes due to your rising SG levels, the
SmartGuard suspend features will be unavailable for 30 minutes after your
basal insulin delivery resumes.
• Manually resume your basal insulin delivery, the SmartGuard suspend features
will be unavailable for 30 minutes after your basal insulin delivery resumes.
If your pump has been suspended for two hours and you have not responded,
basal insulin delivery automatically resumes.
If you respond within 30 minutes of basal insulin delivery being resumed, the
SmartGuard suspend features will be unavailable for a total of 30 minutes. For
example:
• If you respond 10 minutes after your basal insulin delivery resumes, the
SmartGuard suspend features will be unavailable for an additional 20 minutes.
• If you respond 20 minutes after your basal insulin delivery resumes, the
SmartGuard suspend features will be unavailable for an additional 10 minutes.
If you respond 30 minutes to four hours after your basal insulin delivery resumes,
the SmartGuard suspend features will be available immediately.
If you do not respond, the SmartGuard suspend features will be unavailable for
four hours after basal delivery resumes.
Alert on low
The Alert on low feature is automatically turned on when either the Suspend
before low or the Suspend on low feature is turned on.
■
Setting up CGM
Setting up CGM 185