How-To Guide

If your ratio value is outside the range of 5 to 50 grams per unit, a message
appears asking you to confirm your setting.
4. When the Edit Sensitivity screen appears, enter your insulin sensitivity factor.
You can set up to eight different sensitivity factors using different time
segments. The time segments must cover a 24-hour period.
If the value you enter is outside the range of 20 to 100 mg/dL per U, a
message appears asking you to confirm your setting.
5. When the Edit BG Target screen appears, enter your Bolus Wizard BG target
range. You can set up to eight different BG target ranges using different time
segments. The time segments must cover a 24-hour period.
If your Bolus Wizard BG target is outside the range of 90 to 140 mg/dL, a
message appears asking you to confirm your setting.
6. When the Active Insulin Time screen appears, enter your active insulin time
value. The default is four hours.
Bolus
Bolus 93