Cleaning your transmitter
Always refer to your transmitter user guide for instructions on cleaning the
transmitter.
Storing your pump
Storage mode lets you safely place your pump in storage while not in use.
Note: If you place your pump in storage mode, it is important to insert a
new AA battery for 8 to 12 hours every six months to ensure that the
internal battery does not discharge to a deep discharge. A battery that is
deeply discharged may experience decreased performance.
WARNING: After placing your pump in storage mode, do not rely
on active insulin tracked in the pump when making new Bolus
Wizard calculations. Storage mode clears active insulin. Inaccurate
Bolus Wizard calculations could result in inaccurate insulin delivery,
and serious injury.
To place your pump in storage mode:
1. Remove the AA battery from the pump. For details, see Removing the battery,
on page 39.
Note: When you remove the battery, your pump issues an Insert
Battery alarm for 10 minutes or until you place your pump into
storage mode.
2. Press and hold
until your screen turns off.
CAUTION: Never expose the pump to temperatures below -4°F
(-20°C) or above 122°F (50°C) while it is in storage without a
battery. Storing your pump in temperatures outside of this range
can damage your pump.